Drawell GC1290 Gas Chromatography (LCD Touch Screen)
Available in stock
The Drawell GC1290 Gas Chromatography (LCD Touch Screen) is a professional analytical laboratory instrument. Review features and technical specifications, then request configuration, MOQ, lead time, warranty, and a formal quotation.
Scientific China
Store| Product Category | Chromatography > Gas Chromatography |
|---|
Product Enquiry
Description
Drawell GC1290 Gas Chromatography (LCD Touch Screen)
Product overview
Gas Chromatography is a kind of instrument for qualitative and quantitative analysis of multicomponent complex mixture by using chromatographic separation technology and detection technology. Generally, it can be used to analyze the thermal stable organic compounds in soil, such as volatile organic compounds, organic chlorine, organic phosphorus, polycyclic aromatic hydrocarbons, phthalates, etc.
The Drawell GC1290 Gas Chromatography (LCD Touch Screen) supports research, pharmaceutical, environmental, food, industrial, and quality-control laboratories. Scientific China provides model selection, configuration review, export coordination, and formal quotations.
Key features
- PC Control System for instrument monitoring, operation, and method setup.
- 10-ramp Column Oven with fast heating, rapid cooling, and auto back-door opening.
- Built-in Data Acquisition for real-time status monitoring and signal collection.
- Up to 3 Columns & 3 Injectors with independent injector temperature control.
- Flexible injection system : 2 sample injectors with independent temperature control
- High-sensitivity Stable Detector with dual independent analog outputs.
- M6 Software supports GLP and FDA 21 CFR Part 11 compliance.
- Operation by LCD touch screen
- GC1290 is suitable for volatile and semi-volatile organic compounds in environmental, chemical, food, pharmaceutical, petroleum, and laboratory QC applications.
- The column oven supports ambient temperature +7℃ to 400℃, meeting routine GC analysis requirements for different boiling-point compounds.
- The column oven can accommodate up to 3 chromatographic columns, giving laboratories more method flexibility.
- Yes. 2 sample injectors could be installed and operated simultaneously with independent temperature control.
- GC1120 can be configured with different detectors according to application needs, such as FID, TCD, FPD, and NPD/TSD.
- Yes. With a high-sensitivity and stable detector configuration, GC1290 supports reliable detection for low-concentration components.
Applications
- GC1290 gas chromatography is a chromatographic analysis method with gas as the mobile phase, which is mainly used for the separation and analysis of volatile substances. Gas chromatography has become one of the most important separations and analysis methods, and has been widely used in the fields of medicine and health, petrochemical industry, environmental monitoring, biochemistry, and so on.
Technical specifications
| Model | DW-GC1290 | |
|---|---|---|
| Description | EPC(electronic pneumatic control) 7 inch touch screen display | |
| Sample Introduction System | sample injector and evaporation chamber | |
| Column Oven | Temperature range | Ambient temperature +7℃ ~ 400℃ (in 1℃incremen |
| Temperature accuracy | ± 0.02℃ | |
| Cooling time | 350℃〜100℃≤3min, | |
| 400℃ to 50℃ in 8-10 min at 25℃ ambient | ||
| Programmed temperature setting | 0.1℃~40℃/min (in 1℃ increment) | |
| Program ramps | 20 ramps in total (10 ramps available with | |
| control workstation | ||
| Size (L*W*H) | 284*280*241mm(inside)340*345*281mm(outside) | |
| Flame Ionization | Detection limit | ≤8×10- 12 g/s (n-hexadecane) |
| Detector (FID) | Drift | ≤1.5×10- 13 A/h |
| Thermal Conductivity Detector(TCD) | Noise | ≤3×10- 14 A |
| Linearity range | ≥106 | |
| Flame Photometric Detector(FPD) | Detection limit | ≤8×10- 13 g/s(P) ; ≤8×10- 11 g/s(S) |
| Drift | ≤2×10- 11 A/30min | |
| Baseline noise | ≤5×10- 12 A | |
| Nitrogen-Phosphorus Detector(NPD) | Detection limit | ≤5×10- 13 g/s(P) (Malathion); |
| ≤5×10- 12 g/s(N) (Azobenzene) |
Ordering and configuration
Final specifications, included accessories, consumables, software, certifications, power requirements, installation, warranty, and regulatory availability depend on the selected model and destination. Contact Scientific China to confirm the suitable configuration, minimum order quantity, price, lead time, documentation, and international delivery terms before purchase.
Reviews (0)
Only logged in customers who have purchased this product may leave a review.























































Reviews
There are no reviews yet.