Drawell DTR3110-1064 Portable Raman Spectrometer
Available in stock
The Drawell DTR3110-1064 Portable Raman Spectrometer is a professional analytical laboratory instrument. Review features and technical specifications, then request configuration, MOQ, lead time, warranty, and a formal quotation.
Scientific China
Store| Product Category | Spectrophotometer > Raman Spectrometer |
|---|
Product Enquiry
Description
Drawell DTR3110-1064 Portable Raman Spectrometer
Product overview
DTR3110-1064 portable Raman spectrometer is suitable for field operation. The outstanding reliability makes the detection result much more accurate. The excellent low stray light conditions that enable the spectrometer has a wide range of application, especially in public safety, food safety, and pharmaceutical engineering. The multi-function software facilitated the spectral analysis process in application and the remote experiment through internet access makes the test item much easier.
The Drawell DTR3110-1064 Portable Raman Spectrometer supports research, pharmaceutical, environmental, food, industrial, and quality-control laboratories. Scientific China provides model selection, configuration review, export coordination, and formal quotations.
Key features
- High sensitivity 512 pixels InGaAs array; TE-cooled, down to -20℃,
- Ultra-low noise circuit;
- Powerful embedded software;
- Fluorescence background elimination;
- Peak value search and display;
- User-friendly human and machine interface;
- DTR3110-1064 System
- Operating system
- Integration time
- 10ms-600000ms
- Power voltage
- DC 5V(+/-5%)
- Operating Temp
- Operating humidity
Applications
- Biological science
- Material science
- Pharmaceutical engineering
- Forensic analysis
- Agriculture and food safety
- Environmental science
Technical specifications
| DTR3110-1064 System | |
|---|---|
| Interface | USB2.0 |
| Operating system | Windows10 |
| Integration time | 10ms-600000ms |
| Power voltage | DC 5V(+/-5%) |
| Operating Temp | -10〜40℃ |
| Operating humidity | < 95% |
| Dimension(L*W*H) | 26x33x16.5cm |
| Weight | 5.5Kg |
| Reliability | |
| Spectral stability | σ/μ < 0.5%(COT8hours) |
| Temp stability | Spectral shift< 1cm-1 (10-40℃) |
| Variation of intensity (in5~40℃) | <±5% |
| Optical parameters | |
| 一1 Spectral range (cm) | 200-2600 |
| -1 resolution (cm) | 12-15 |
| SNR | >3000:1 |
| Slit size | 50um, 25um optional |
| Optical system | f/4 C-T crossed optical path |
| Max Quantum Efficiency | >90% |
| Focal length | 112mm for incidence and output |
| Detector | |
| Item | High sensitivity512pixels InGaAs array |
| Cooled down to | -20℃ |
| Detecting range | 900-1700nm |
| Effective pixels | 512 |
| Dynamic range | 14,000: 1 |
| Pixel size | 25×500 μm |
| Exciting Laser | |
| Central wavelength | 1064nm (+/-1nm) |
| Semi-peak width | 0.1 nm |
| Power output | ≥500mW |
| Power stability | σ/μ <±0.2% |
| Raman probe | |
| Operating distance | 6mm |
| Rayleigh scattering resistance | OD>8 |
| Numerical Aperture | 0.3 |
| Aperture | 7mm |
Ordering and configuration
Final specifications, included accessories, consumables, software, certifications, power requirements, installation, warranty, and regulatory availability depend on the selected model and destination. Contact Scientific China to confirm the suitable configuration, minimum order quantity, price, lead time, documentation, and international delivery terms before purchase.
Reviews (0)
Only logged in customers who have purchased this product may leave a review.




















Reviews